Alarms
Active and/or latched alarms
Selection
Result
Select
once
●
Pauses all active audio alarms for 2 minutes.
●
Removes all latched alarms.
●
Deactivates some technical alarms.
Second selection of
during
the 2 minute pause
●
Starts a 2 or 5 minute audio pause period for all alarms except the
specified breakthrough alarms. The 2 or 5 minute duration is a care unit
setting and password protected.
●
Removes all new latched alarms.
Some technical alarms may also be deactivated with this selection.
Select
once during audio
pause
●
Ends the audio pause period.
●
Restores all acknowledged and silenced alarms if the alarm condition
still exists.
No active or latched alarms
Selection
Result
Select
once
●
Starts a 2 or 5 minute audio pause period for all alarms except the
specified breakthrough alarms.
Select
once during audio
pause
●
Ends the audio pause period.
●
Restores all acknowledged and silenced alarms if the alarm condition
still exists.
Pausing alarms for 5 minutes
You can pause audible alarms with the pause audio key for 2 or 5 minutes according
to the care unit settings. You can also pause all alarms for 5 minutes through the
Alarms Setup
menu.
1.
Select
Alarm Setup
from the monitor’s main menu.
2.
Select the
Audible & Visual
tab.
3.
Select
Pause All Audio for 5 min
. This will pause all alarms, including the
breakthrough alarms, except
FiO2 low
,
EtO2 low<18%
,
FiN2O low>82%
, and
Ppeak high
. It also removes latched alarms.
Activating all audible alarms
If necessary, you can activate all paused audible alarms before the 2 or 5 minute
pause expires.
1.
Select
Alarm Setup
from the monitor’s main menu.
2.
Select the
Audible & Visual
tab.
3.
Select
Activate All Audible Alarms
.
138
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...