Starting and ending monitoring
End of standby
The monitor ends the standby automatically when any of the following conditions
occur:
●
Vitals signs are still present after
Prepare for Standby
has been selected and the
audio pause time expires.
●
Any vital signs are detected as active.
●
User input is received: a keyboard key is pressed, Trim Knob is pressed or rotated,
primary mouse button is pressed, touchscreen is pressed.
●
A PDM or TRAM is connected.
About continuing monitoring
Patient information and data is stored in the PDM and TRAM. The
Connecting
Measurement
message appears on the monitor when the PDM or TRAM is first
connected. Vital sign monitoring is not available during this initialization time. At
the conclusion of the initialization, the stored patient information and data can be
transferred to the monitor from the acquisition module. While patient information and
data is being transferred, the
Loading from PDM
or
Loading from Tram
message
appears on the monitor. The type of patient information and data that can be
transferred includes the following:
●
Patient demographic data
●
Patient trends and alarm histories; alarm histories are converted to snapshots
during the transfer
●
All zeroed invasive pressure site labels
●
All invasive pressure transducer zero values for channels 1 to 4
●
Latest values and timestamps for C.O., PCWP, and NIBP measurements
●
NIBP cuff size
●
NIBP auto cycle on/off information
●
NIBP cycle time information if the NIBP auto cycle is on
The actions taken by the monitor and the menus that appear depend on whether the
monitor has an active patient case/admitted patient or not.
How to continue monitoring when a case is not
active/patient is discharged
When you connect a PDM or TRAM that is actively measuring vital signs from ECG,
invasive pressures or SpO
2
, the patient information and data is automatically loaded
from the PDM or TRAM to the monitor and a case is automatically started/patient is
admitted.
When you connect a PDM or TRAM that is not actively measuring vital signs from
ECG, invasive pressure or SpO
2
but contains patient information, the
Continue
menu
appears. The
Continue
menu has two informative fields on the top: the
Patient in the
monitor
and the
Patient in the PDM
or
Patient in the Tram
. These fields show the text
No patient identification data available
if there is no active case/admitted patient.
Select one of the following to continue monitoring:
●
Load PDM Data
/
Load Tram Data
: This selection loads patient information and
data from the module.
2062971-001
CARESCAPE Modular Monitors
123
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...