2048723-002C
CARESCAPE V100 Vital Signs Monitor
3-15
Installation:
Host communication connector
Host communication connector
CAUTION
Auxiliary equipment connected to the CARESCAPE V100 vital
signs monitor will result in the formation of an electromedical
system and thus, must comply with the requirements of IEC
60601-1-1. All host port signals are NON-ISOLATED and should
be connected to equipment conforming to IEC-60601-1 or
configured to comply with IEC 60601-1-1
only
.
Where isolation of data communication is required, use one of the following:
ILC-1926 Isolated Line Converter (GE part number 001926) along with the
Host Comm Cable Assemblies (GE part numbers 683235 and 683242).
USB Cable Kit (GE part number 2040229-001).
If external alarm control is required, the Isolated Remote Alarm Cable (GE part
number 487208CR) should
always
be used.
When a high-priority alarm condition is displayed on the monitor, the remote
alarm signal becomes active within 0.5 seconds. The active state of the alarm
signal is an open circuit. In the inactive state the alarm signal is connected to
ground. Refer to the Information Sheet included with the Isolated Remote Alarm
Cable for operational details.
NOTE
When using the Isolated Remote Alarm Cable, the visual and audible alarms
of the monitor should still be considered the primary alarm delivery method.
The Remote Alarm connection should be considered a secondary method.
Содержание CareScape V100
Страница 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Страница 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Страница 42: ...2 18 CARESCAPE V100 Vital Signs Monitor 2048723 002C Equipment overview Functional description For your notes ...
Страница 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Страница 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Страница 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Страница 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Страница 98: ...5 8 CARESCAPE V100 Vital Signs Monitor 2048723 002C Troubleshooting Alarm code interpretation For your notes ...
Страница 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Страница 127: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 29 Parts lists and drawings Field replaceable units FRUs ...
Страница 128: ...6 30 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 129: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 31 Parts lists and drawings Field replaceable units FRUs ...
Страница 130: ...6 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 131: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 33 Parts lists and drawings Field replaceable units FRUs ...
Страница 132: ...6 34 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 133: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 35 Parts lists and drawings Field replaceable units FRUs ...
Страница 145: ...2048723 002C CARESCAPE V100 Vital Signs Monitor A 1 A Technical specifications and default settings ...
Страница 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Страница 177: ......