CARESCAPE Monitor B850
76
2106778-001
Citrix Timeout:
The number of minutes that a Citrix session will remain open after the
Citrix Window is closed. Keeping the session alive allows users to
quickly return to their last session but locks a Citrix license.
Citrix Startup Time:
Estimated time, in seconds for connecting Citrix. During this time, the
Citrix Window is displayed as blank, grey rectangle.
7.7 Configuring MUSE/12SL
NOTE: MUSE viewing is a licensed software feature. You can configure the MUSE/12SL settings
independent of the license status, but actual viewing of MUSE reports requires that the
MuseView license is enabled.
Settings to send 12SL data
1.
Log in to Webmin.
2.
Select
Configuration
>
MUSE/12SL
.
3.
In the
MUSE/12SL
window, enter the applicable information in the following fields:
4.
Select
Save
.
These settings take effect immediately after they are submitted.
Settings to view 12SL data
1.
Log in to Webmin.
2.
Select
Configuration
>
MUSE/12SL
.
3.
In the
MUSE/12SL
window, enter the following:
4.
Select
Save
.
These settings take effect immediately after they are submitted.
- Location ID
Identifies the location ID number (within the range 0 to 599)
associated with the patient monitor for searching the MUSE
system.
- Site Number
Identifies the site number (within the range 1 to 254) associated
with the patient monitor for searching the MUSE system.
- MUSE Web Username
Username used to authenticate with the MUSE
Web to access 12-lead reports.
- MUSE Web Password
Password used to authenticate with the MUSE
Web to access 12-lead reports.
- Confirm Password
The password and the confirmed password must
be the same.
- MUSE Web URL
Used to locate the MUSE Web system to access
12-lead reports. Enter the URL in a valid format.
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......