CARESCAPE Monitor B850
74
2106778-001
7.5.2 Deleting a printer
NOTE: Before deleting a laser printer, check
Monitor Setup > Printing > Devices > Status.
If a
printout is assigned to the printer to be deleted, redirect the printout to another valid printer.
1.
Log in to Webmin.
2.
Select
Configuration
>
Printers
.
3.
In the
Sub-Modules for Printers
menu, select
Delete Printer
.
4.
In the
Delete Printer
window, select the printer to delete.
5.
Select
Save
.
The change will take effect immediately.
7.5.3 Printing a test page
1.
Log in to Webmin.
2.
Select
Configuration
>
Printers
.
3.
In the
Sub-Modules for Printers
menu, select
Print Test Page
.
4.
In the
Print Test Page
window, select the printer.
5.
Select
Save
.
7.6 Configuring Citrix
Configure the Citrix connection for iPanel software.
NOTE: GE is not responsible for installing or configuring the Citrix server.
NOTE: Contact hospital's IT Administrator or biomedical department to obtain values to the
following fields.
NOTE: Server address, initial program and Citrix session timeout are mandatory fields.
1.
Log in to Webmin.
2.
Select
Configuration
>
Citrix
.
3.
In the
Citrix Configuration
window, enter the applicable information in the following
fields:
Field name
Description and valid choices
Server Version
Select the supported server version: 4.0, 4.5,
5.0, 6.0, 6.5.
Server 1 to Server 4 Addresses
Blank, server IP address or server hostname,
up to 255 character long.
Initial Program
Initial Citrix- published application that will be
launched, for example #MUSE. This field is
mandatory, if the server address has been
entered. The maximum length is 128
characters.
Citrix Session Timeout (in Minutes)
Valid values are between 0-99. Keeping the
session alive allows users to quickly return to
their last session, but locks a Citrix license.
Selecting "0" disables the timeout feature.
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......