Configuration
73
2106778-001
NOTE: The same unit name must be set for all the patient monitors and CARESCAPE Central
Stations (CSCS) that are connected to the same care unit in the MC Network. Bed name shall be
unique to each patient monitor in the same care unit.
NOTE: Unit and bed name selections are not available if network selection is S/5.
1.
Log in to Webmin.
2.
Select
Configuration
>
Unit and Bed Name
.
3.
In the
Unit and Bed Name Configuration
window, view or set the unit name and bed
name for the device.
NOTE: Use only capital letters A - Z, numbers, dash (-), asterisk (*) and space ( ). The unit name
may be up to seven characters long and bed name up to five characters long.
4.
Select
Save
.
The unit and bed name configuration takes effect immediately.
7.5 Configuring printers
You can configure the patient monitor to print to up to 12 laser printers connected on the IX
Network. Use the Webmin sub-modules to install or delete a printer and to print a test page.
NOTE: Ensure that you have the host name(s) or IP address(es) for all the connected IX printers
available. Configure the IX printers according to this information.
NOTE: Laser printers that are installed on central stations do not need to be configured using
Webmin. Refer to the user’s manual for details on configuring the printer.
NOTE: Refer to the patient monitor's supplemental information manual for the list of supported
devices.
7.5.1 Installing a printer
1.
Log in to Webmin.
2.
Select
Configuration
>
Printers
.
3.
In the
Sub-Modules for Printers
menu, select
Install Printer
.
4.
Below
Printer Configuration Information
in the
Install Printer
window, provide the
following information:
a.
Select either the
Hostname
or
IP Address
radio button, as applicable.
b.
In the
Hostname
or
IP Address
field, enter the printer
Hostname
or
IP Address
.
c.
In the
Printer Name
field, enter the
Printer Name
.
d.
Select
Yes
from the drop-down list next to
Test Page
.
5.
Select
Save
.
6.
From the patient monitor, select
Monitor Setup
>
Printing
.
7.
Select the
Devices
>
Setup
.
8.
In the
Printout
menu, select what to print out (for example,
Waveforms
,
Alarm
Waveforms
,
Numeric Trends
,
Reports
).
9.
Below
Location
, select the radio button next to
Network
.
10. From the drop-down list next to
Network Device
, select the desired printer.
The change will take effect immediately.
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......