1.
Check that a network symbol
is displayed in the upper right corner of the
screen.
2.
Select
Data & Pages – Other Patients
and select a patient from the list.
3.
Select
View
and verify that a window with parameters from another patient displays on
the left side of the screen.
4.
Select
Close View
to close the window.
8.3.10 IX printers
Perform the following test only if the patient monitor is connected to a printer in the IX Network
and you did not print a test page while you configured the IX printer.
1.
Log in to Webmin.
2.
Select
Configuration
>
Printers > Print Test Page
.
3.
Select the IX Printer to test.
4.
Select
Save
.
5.
Verify that the test page was printed to the selected printer.
6.
Repeat steps 3 to 5 for all connected IX printers.
8.3.11 iPanel connection
Perform the following test only if the Citrix / iPanel is configured and in use.
1.
Select
Data & Pages > iPanel
and verify that the initial program (configured in Webmin) is
launched correctly.
2.
Select
to exit the Citrix thin client.
8.3.12 InSite RSvP
Perform the following test only if remote service is configured and enabled, and you did not test
the connectivity already while you enabled it.
1.
Log in to Webmin
2.
Select
Configuration > Remote Service
.
3.
In the
Sub-Modules for Remote Service
menu, select
Control
.
4.
In the
Remote Service Control
window under
Test Connectivity
, select
Test
to test the
connectivity to the remote service server.
5.
Verify that
Connection to Enterprise URL
and
Connection to Tunnel URL
both "Pass" the
test.
If one of the connection tests fail, check the failure description given and fix the connectivity
issue. Re-test the connectivity.
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......