Unity Network ID Interface Device
9.
Visually ensure that the device is clean. If visible soil remains, repeat cleaning
procedure until the device is thoroughly clean.
10. Reconnect the equipment to the power supply.
11. Reconnect all cables.
12. Turn on the power to the equipment.
Other manufacturer cleaning instructions
The information provided in this chapter does not supersede any instructions for use
provided by the manufacturer or provided with a device, device component, supply,
or accessory.
For cleaning, disinfection, and care information for devices, device components,
supplies, and accessories made by manufacturers other than GE, see the applicable
instructions for use provided by the manufacturer.
Disinfection
Disinfection safety precautions
Disinfection cautions
CAUTION
IMMERSION. Never immerse Unity Network ID interface device
components in cleaning fluids or disinfecting agents.
Disinfection notices
NOTICE
COLOR SHIFT. Some plastics may experience as light change
in color over time due to effects of disinfecting agents. This
effect is cosmetic and not associated with product safety or
normal function.
NOTICE
SPOTTING, STAINING, CORROSION. Metal components
including fasteners may corrode or become stained or
spotted in appearance from continuous exposure to chlorine
compounds like bleach. These effects are not associated with
product safety or normal function.
Disinfection points to note
●
Always clean before disinfecting.
●
Always dilute cleaning and disinfectant agents according to their manufacturer’s
instructions. Always consider your hospital guidelines as well.
●
Use only the permitted substances.
●
Your hospital guidelines permitting, all cleaning and disinfection activities can
be carried out at the bedside.
●
Before drying a surface after wiping it, wait the minimum time required according
to the substance manufacturer’s instructions.
●
Visually ensure that no substance residue remains on the equipment.
2110145-001
Cleaning and disinfecting
61
Содержание CARESCAPE B105
Страница 8: ...8 Cleaning and disinfecting 2110145 001 ...
Страница 30: ...ApexPro and CARESCAPE Telemetry 30 Cleaning and disinfecting 2110145 001 ...
Страница 48: ...CARESCAPE B450 Monitor 48 Cleaning and disinfecting 2110145 001 ...
Страница 54: ...CARESCAPE B650 Monitor 54 Cleaning and disinfecting 2110145 001 ...
Страница 71: ...content ...