Alarms
4-7
Adjusting alarm limits
Adjusting limits
1.
Press the
Alarms Setup
side panel key.
2.
Select
Adjust Limits
.
3.
Turn the ComWheel to highlight the measurement. If the desired measurement is not
displayed in the window, select
Next Page
.
4.
Push the ComWheel to open an adjustment window.
5.
Turn the ComWheel to change the limits and accept them by pushing it. Move between
selections by turning the ComWheel.
6.
To return to the
Alarms Setup
menu to select more measurements, push the ComWheel
until the cursor is in the adjustment menu, then select
Previous Menu
or
Alarms Setup
.
7.
Press the
Normal Screen
key to return to normal monitoring view.
You can enter the alarm limit adjustment window also through the measurement menus.
When you are in an XX parameter menu, select
XX Alarm
or enter
XX Setup
and then select
XX Alarm
.
NOTES:
•
ST high/low alarms will reach the white (low priority) level only.
•
In NIBP measurement, the alarm limits change automatically according to the cuff hose
type used.
•
If the monitor is connected to the network, the alarm limits can also be adjusted using the
Central if this feature has been enabled in its configuration. If alarm limits are adjusted
using the Central, the message 'Alarm settings changed from Central' is displayed in the
bedside monitor.
Choosing automatic limits
To activate automatic patient-specific alarm limits enabling close patient control, select
Auto
Limits
in the
Alarms Setup
menu. Limits are then calculated from the displayed patient reading
at the point of time when auto limits are selected.
Returning to default limits
Select
Default Limits
to set the alarms to the default alarm limits. The factory default alarm
limits are listed in the "Default Configuration Worksheet."
Содержание b30
Страница 2: ......
Страница 8: ......
Страница 10: ...Table of contents ii...
Страница 12: ...Table of contents ii...
Страница 16: ...B30 Patient Monitor 4...
Страница 18: ...Table of contents ii...
Страница 26: ...Table of contents ii...
Страница 62: ...B30 Patient Monitor 2 36...
Страница 64: ...Table of contents ii...
Страница 72: ...Table of contents ii...
Страница 90: ...Table of contents ii...
Страница 106: ...Table of contents ii...
Страница 118: ...B30 Patient Monitor 6 12...
Страница 120: ...Table of contents ii...
Страница 124: ...B30 Patient Monitor 7 4...
Страница 126: ...Table of contents ii...
Страница 136: ...Table of contents ii...
Страница 144: ...B30 Patient Monitor 9 8...
Страница 146: ...Table of contents ii...
Страница 156: ...Table of contents ii...
Страница 176: ...Table of contents ii...
Страница 184: ...B30 Patient Monitor 12 8...
Страница 186: ...Table of contents ii...
Страница 192: ...Table of contents ii...
Страница 202: ...Table of contents ii...
Страница 212: ...B30 Patient Monitor 15 10...
Страница 214: ...Table of contents ii...
Страница 224: ...B30 Patient Monitor 16 10...
Страница 226: ...Table of contents ii...
Страница 236: ...B30 Patient Monitor 17 10...
Страница 241: ......