B30 Patient Monitor
3-32
Document no
.
2044677-001
5
Adjustments and calibrations
NOTE: Use only properly maintained, calibrated and traceable measurement equipment for
the specified calibrations and adjustments to ensure accuracy.
5.1 NIBP calibrations
The electronics of the NIBP pressure measurement is calibrated at the factory. The processor
automatically maintains the zeroing pressure. If the zero point of the pressure transducer drifts
more than specified, an error message is given and the NIBP board should be recalibrated or
replaced.
Recalibrate the NIBP measurement once a year. The checking and recalibration can be done in
the NIBP service menu.
The calibration of the primary pressure channel can also be checked from the NIBP setup
menu (
NIBP - NIBP Setup - Calibration Check
). In this case, the auto zeroing is performed at
start - remove the hose before entering to ensure atmospheric pressure to the pressure
transducers - the primary pressure is displayed. The zero-offset value should then be zero.
Check the intake air filter as part of the calibration check. Change the filter if it is visibly dirty.
Calibration check
1.
Enter
Calibration
menu:
Monitor Setup -
Install/Service
(password 16-4-34) -
Service
(26-23-8) -
Parameters -
NIBP - Calibrations
2.
Select
Calibration Check
and push the ComWheel.
3.
Connect an external precision manometer to the module.
4.
Pump the following pressures to manometer and check the difference between the
manometer and monitor pressure display (The zeroing offset is automatically subtracted
from the pressure readings).
PR
ES
TN_cali
b
_menu.vsd
Содержание b30
Страница 2: ......
Страница 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Страница 13: ...1 Introduction...
Страница 14: ......
Страница 61: ...2 Installation...
Страница 69: ...3 Maintenance...
Страница 70: ......
Страница 105: ...4 Troubleshooting...
Страница 106: ......
Страница 125: ...5 Service Menu...
Страница 126: ......
Страница 151: ...Parameters 5 25 Document no 2044677 001 4 4 4 NIBP Buttons Leds Buttons Leds Data Not in use in B30 Monitor...
Страница 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Страница 159: ...6 Field replaceable unit...
Страница 160: ......
Страница 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Страница 183: ...Frame disassembly 6 23 Document no 2044677 001...
Страница 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Страница 191: ...7 Technical specification...
Страница 192: ......
Страница 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Страница 199: ...8 N F C REC Module...
Страница 200: ......
Страница 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Страница 244: ...B30 Patient Monitor C 2 Document no 2044677 001 Notes Signature Functional Inspection OK N A Fail...
Страница 251: ......