ECG specifications
Leads available
3-lead configuration: I, II, III
5-lead configuration: I, II, III, aVR, aVL,
aVF and VA
QRS detection range
0.5 to 5mV
QRS detection width (Q to
S)
40 to 120 ms
Defibrillation protection
5000 V, 360 J
Recovery time
<5 s
Input impedance
Common mode > 10 M
Ω
@ 50/60 Hz
Differential > 2.5 M
Ω
from 0.67 to 40
Hz
Common mode rejection
90 dB minimum at 50 Hz
Tall T wave rejection
>1.4 mV
ECG leads off detection
Active patient electrode: <30 nA
Reference electrode: <300 nA
Filter modes
50/60 Hz
Monitoring filter
0.5 to 40 Hz
ST filter
0.05 to 40 Hz
Dagnostic filter
0.05 to 150 Hz
Heart rate
Measurement range
30 to 300 bpm
Measurement accuracy
±5 % or ±5 bpm, whichever is greater
resolution
1 bpm
Heart rate response time (IEC 60601-2-27 201.7.9.2.9.101 b) 5)
Step increase from 80 to
120 bpm
average 6.9 s (6.5 to 7.5 s)
Step decrease from 80 to
40 bpm
average 8.2 s (7.6 to 10.0 s)
The heart rate calculation operates with irregular rhythms of IEC
60601-2-27 201.7.9.2.9.101 b) 4), the heart rate after a 20 second
stabilization period is:
Figure 3a
80 bpm
Figure 3b
59 bpm
Figure 3c
122 bpm
Figure 3d
117 bpm
Heart rate averaging computation (IEC 60601-2-27 201.7.9.2.9.101
b) 3)): Average of 10 second median values
The average time and time range ( ) to alarm (V Fib or V Tach) for
tachycardia waveform are as follows (IEC 60601-2-27
201.7.9.2.9.101 b) 6))
Figure 4a halved
amplitude:
9.9 s (8.4 to 11.5 s)
Figure 4a
normal amplitude:
7.1 s (5.8 to 8.2 s)
Figure 4a
doubled amplitude:
4.4 s (4.2 to 4.6 s)