The
Pressure Hold
,
Hold Time
, and
PEEP on Exit
settings can be
preset by the Super User. These settings can be changed by the
user before starting the procedure.
PEEP on Exit
shows if it is set to
Yes
by the Super User.
Using vital capacity
1.
Select
Procedures
.
2.
Select
Start Vital Capacity
.
One pressure breath is given at the set pressure.
The pressure is held for the set time.
PEEP is set to the
PEEP on Exit
setting.
3.
Select
Stop Vital Capacity
at any time to stop the procedure.
4.
Select
Close
.
If the procedure is stopped before completion, the
PEEP on Exit
setting is not used.
Changing vital capacity settings
1.
Select
Procedures
.
2.
Select the setting to change and make the change.
•
Set
Pressure Hold
to between 20 and 60 cmH2O.
•
Set
Hold Time
to between 10 and 40 seconds.
•
Set
PEEP on Exit
to Off or between 4 and 30 cmH2O.
3.
Select
Start Vital Capacity
.
4.
Select
Close
.
Cycling
Use the Cycling procedure to deliver pressure breaths through a
series of ventilation steps. The Cycling procedure provides a flexible
way to deliver pressure breaths during ventilation without making
multiple ventilator setting changes. Up to seven preset steps with
multiple breaths are available.
Each procedure defaults steps and ventilation settings which can be
preset by the Super User. The ventilation settings of each step can
be changed by the user before starting a procedure.
There is a limited amount of gas in the bellows. No additional gas
enters the bellows during the cycling procedure. Increase fresh gas
flow to avoid bellows collapse.
Using cycling
1.
Select
Procedures
.
Note
Note
Note
3 Operation
2067226-001
3-29
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...