Airway module specifications
Use only airway modules that have anesthetic agent monitoring and
O2 monitoring on this system. The following modules can be used on
this system:
•
E-series: E-CAiO, E-CAiOV, E-CAiOVX (software version 4.5
and above)
•
M-series: M-CAiO, M-CAiOV, M-CAiOVX (software version 3.2
and above)
•
CARESCAPE series: E-sCAiO and E-sCAiOV
Gas specifications for E-, M-series modules
.
E- and M-series
Airway humidity
0 to 100% condensing
Sampling delay
2.5 seconds typical with a 3 m sampling line
Total system response time
2.9 seconds typical with a 3 m sampling line, including sampling delay and rise time
Warm-up time
2 minutes for operation with CO2, O2, and N2O
5 minutes for operation of anesthetic agents
30 minutes for full specifications
Respiration rate
4 to 60 breaths/min
Diversion flow
200 ml/min +/- 20 ml/min
Airway pressure
-20 cmH2O to 100 cmH2O
.
E- and M-series airway modules: Accuracy under different conditions
Ambient temperature: 10 to 40°C
Ambient pressure: 500 to 800 mmHg, +/- 50 mmHg of calibration
Ambient humidity: 10 to 98% RH, +/- 20% RH of calibration
Automatic compensation for ambient pressure
Full module accuracy for Respiration rate of 4 to 35 breaths/min
During warm-up 10 to 30 minutes, under normal
conditions
During warm-up 2 to 10 minutes (anesthetic
agents 5 to 10 minutes), under normal
conditions
CO2
+/- (0.3 vol% + 4% of reading)
+/- (0.4 vol% + 7% of reading)
O2
+/- (2 vol% + 2% of reading)
+/- (3 vol% + 3% of reading)
N2O
+/- (3 vol% + 3% of reading)
+/- (3 vol% + 5% of reading)
Hal, Enf, Iso,
Sev, Des
+/- (0.2 vol% + 10% of reading)
+/- (0.3 vol% + 10% of reading)
11 Specifications and theory of operation
2067226-001
11-19
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...