Calibrating the airway module
Calibrate airway modules once every six months or whenever there
are indications of errors in the gas readings. Use a manufacturer
approved calibration gas and regulator to calibrate the modules. See
the "
" section for the stock numbers of the calibration gas and
regulator.
Only use manufacturer approved calibration gas. Do not
use any other calibration gases or the calibration will not
succeed. Dispose of calibration gas containers in
accordance with local environmental procedures.
The
Calibration
menu is not available during
Checkout
or during a
case.
During gas calibration, % units are used for CO2 regardless of
selected measuring units.
1.
Turn on the power. Let the module warm up for 30 minutes
before starting calibration.
2.
Attach the regulator to the calibration gas cylinder.
3.
Attach a new sampling line to the water trap. Connect the loose
end of the sampling line to the regulator on the calibration gas
cylinder.
4.
Select the
System Setup
button.
5.
Select
Calibration
.
6.
Select
Airway Gas
.
7.
Wait until ‘Feed Gas’ appears.
8.
For regulators with a numbered gauge, open the regulator until
the gauge reads between 5 to 7 psi (34 to 48 kPa). For
regulators with a non-numbered gauge, open the valve.
9.
Feed the calibration gas until ‘Adjust’ appears.
•
If an error occurs during calibration or if no gas is fed,
‘Calibration error’ appears. Push the ComWheel to perform
a new calibration.
•
Do not close the regulator until all the adjustments have
been made.
10. If a gas needs adjustment, select the gas to be adjusted. Use
the ComWheel to change the value until it matches the
calibration gas cylinder value. Push the ComWheel to confirm
the change.
If a gas does not need adjustment, select the gas. Push the
ComWheel to confirm the value.
11. Close the regulator.
WARNING
Note
Aisys CS²
6-8
2067226-001
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...