Full test
The
Full Test
or the individual tests must be performed at least once
within every 24-hour period.
Perform the
Full Test
at the start of each day. The full test runs
automatically and beeps to indicate when it is finished or if
interaction is required.
The
Full Test
does the following tests:
Vent and Gas
,
Circuit Leak
,
and
Circuit O2 Cell
(if circuit O2 cell is present). When one of the
tests is completed, the next test begins.
1.
From the
Checkout
menu, select
Full Test
and follow the
instructions.
2.
If a test fails, follow the instructions to perform a retest or accept
the results.
3.
When the
Full Test
is completed, start a case.
In case of a patient emergency, the
Full Test
may be bypassed by
selecting
Start Case
. The general message ‘Please Do Checkout’ is
displayed if a
Full Test
or all of the individual tests are not completed
with passing results within 24 hours.
Vent and gas
The
Vent and Gas
test checks the agent delivery, airway module,
Bag/Vent switch, proper gas supply pressures, ventilator operation
and leak, battery and electrical power, circuit compliance, and flow
control operation. This is a two-step test.
To run this test, follow the on-screen instructions. When the test
passes, the next test starts.
Any cassette may be used during the test. Use a non-desflurane
cassette during the test to check the full functionality of the internal
electronic control unit.
Circuit leak
The
Circuit Leak
test checks the Bag/Vent switch, proper gas
supply pressures, airway pressure measurement transducer, APL
valve, and manual circuit leak.
To run this test, follow the on-screen instructions. When the test
passes, the next test starts.
Note
Note
Aisys CS²
5-6
2067226-001
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...