D-2
ApexPro™
2001989-200E
Troubleshooting
ECG
Arrhythmia troubleshooting
Problem: Inaccurate heart rate and/or false asystole
Solution:
Check ECG signal from patient:
1.
Check/adjust lead placement.
2.
Check/perform skin preparation.
3.
Check/replace electrodes.
Check amplitude of ECG waveform:
1.
Click on the patient’s
ECG
tab.
2.
Click on all ECG leads in the Display section and check for 0.5 mV amplitude at
normal (1X) size. (At least 0.5 mV amplitude is required for QRS detection.) For
borderline signals, validate on a graph.
3.
If amplitudes are low, electrodes may need to be repositioned or replaced.
Relearn arrhythmia:
1.
Click on the patient’s
ECG
tab.
2.
Click on the
Relearn
button.
IF PROBLEM CONTINUES: Change to
Single-Lead
ECG detection and processing:
1.
Click on the patient’s
ECG
tab.
2.
Click on
Single-Lead
in the
Lead Analysis
section.
3.
Click on the ECG leads in the Display section and change top ECG waveform to
display the lead with the greatest amplitude. (At least 0.5 mV amplitude is
required for QRS detection.)
Problem: False ventricular calls
Solution:
Check ECG signal from patient: (V leads may exhibit polarity changes
which may occasionally cause an inaccurate call.)
1.
Check/adjust lead placement.
2.
Check/perform skin preparation.
3.
Check/replace electrodes. (If V lead is a problem, move the V lead to another V
position.)
4.
Relearn ECG:
Click on the patient’s
ECG
tab.
Click on the
Relearn
button.
IF PROBLEM CONTINUES:
1.
Remove the V lead(s).
Содержание ApexPro
Страница 13: ...2001989 200E ApexPro 1 1 1 Introduction ...
Страница 31: ...2001989 200E ApexPro 2 1 2 Equipment overview ...
Страница 36: ...2 6 ApexPro 2001989 200E Equipment overview ...
Страница 37: ...2001989 200E ApexPro 3 1 3 Equipment setup ...
Страница 50: ...3 14 ApexPro 2001989 200E Equipment setup ...
Страница 51: ...2001989 200E ApexPro 4 1 4 System setup ...
Страница 69: ...2001989 200E ApexPro 5 1 5 Alarms ...
Страница 86: ...5 18 ApexPro 2001989 200E Alarms ...
Страница 87: ...2001989 200E ApexPro 6 1 6 Managing patients ...
Страница 109: ...2001989 200E ApexPro 7 1 7 Viewing real time patient data ...
Страница 144: ...7 36 ApexPro 2001989 200E Viewing real time patient data ...
Страница 145: ...2001989 200E ApexPro 8 1 8 Viewing stored patient data ...
Страница 166: ...8 22 ApexPro 2001989 200E Viewing stored patient data ...
Страница 167: ...2001989 200E ApexPro 9 1 9 Printing ...
Страница 175: ...2001989 200E ApexPro A 1 A Abbreviations and symbols ...
Страница 183: ...2001989 200E ApexPro B 1 B Customized defaults worksheet ...
Страница 186: ...B 4 ApexPro 2001989 200E Customized defaults worksheet ...
Страница 187: ...2001989 200E ApexPro C 1 C Maintenance ...
Страница 194: ...C 8 ApexPro 2001989 200E Maintenance ...
Страница 195: ...2001989 200E ApexPro D 1 D Troubleshooting ...
Страница 206: ...D 12 ApexPro 2001989 200E Troubleshooting ...
Страница 207: ...2001989 200E ApexPro E 1 E Technical specifications ...
Страница 214: ...E 8 ApexPro 2001989 200E Technical specifications ...
Страница 217: ......