7-20
ApexPro™
2001989-351A
Checkout
A CHECK PROBE message appears on the CIC Pro center.
Nonin Xpod
This test verifies the functionality of the Nonin Xpod oximeter.
1.
Record the oximeter’s serial number.
2.
Connect the oximeter to the transmitter.
3.
Admit the transmitter to an ApexPro telemetry system.
a.
Make sure that the SpO2 parameter box comes up on the CIC Pro center.
b.
If the oximeter fails, discontinue the test.
c.
Return the oximeter after double-checking the test setup.
4.
Connect the oximeter to the pulse oximeter simulator (pn 408610-001)using the
Nonin simulator cable adapter (pn 420970-901).
5.
Set the simulator’s selector switch to “Nellcor” and check the heart rate accuracy.
Accuracy specification for 18 – 321 BPM is ± 3 digits with no motion, and for
40-240 BPM is ± 5 digits for high motion.
6.
Adjust the RATE (BPM) switch to vary the heart rate.
a.
Set the RATE to 70 BPM.
Accuracy at 70 BPM is +/–2.
b.
Repeat for 100 BPM (+/–3) and 160 BPM (+/–5).
7.
Set the simulator to 68.4% (use the white Nellcor numbers).
Accuracy at 68% is +/–3.
Repeat for 90.6% (+/–3), 96% (+/–3), and 99% (+/–3).
8.
Disconnect the oximeter from the simulator.
A PROBE message appears on the CIC Pro center.
Accutracker DX NIBP operational tests
GE recommends performing the Accutracker DX NIBP operational tests when you
receive the Accutracker and every 12 months thereafter. If the Accutracker fails any
test, return the unit to GE Service and Supplies.
Transmitter
Oximeter
Adapter
Simulator
Содержание ApexPro T14
Страница 9: ...2001989 351A ApexPro vii ...
Страница 10: ...viii ApexPro 2001989 351A ...
Страница 11: ...2001989 351A ApexPro 1 1 1 Introduction ...
Страница 20: ...1 10 ApexPro 2001989 351A Introduction ...
Страница 21: ...2001989 351A ApexPro 2 1 2 Equipment Overview ...
Страница 41: ...2001989 351A ApexPro 3 1 3 Installation and configuration ...
Страница 62: ...3 22 ApexPro 2001989 351A Installation and configuration ...
Страница 63: ...2001989 351A ApexPro 4 1 4 Maintenance ...
Страница 84: ...4 22 ApexPro 2001989 351A Maintenance ...
Страница 85: ...2001989 351A ApexPro 5 1 5 Troubleshooting ...
Страница 109: ...2001989 351A ApexPro 6 1 6 Replaceable parts ...
Страница 138: ...6 30 ApexPro 2001989 351A Replaceable parts Quad receiver module ...
Страница 139: ...2001989 351A ApexPro 7 1 7 Checkout ...
Страница 162: ...7 24 ApexPro 2001989 351A Checkout Unit Serial Number Institution Name Date Maintenance Repair Technician ...
Страница 163: ...2001989 351A ApexPro A 1 A Technical specifications ...
Страница 179: ...2001989 351A ApexPro B 1 B Electromagnetic compatibility ...
Страница 190: ...B 12 ApexPro 2001989 351A Electromagnetic compatibility ...
Страница 191: ...2001989 351A ApexPro 1 1 ...