5 Setups, alignments and adjustments
64
GE Healthcare
32551-IMG rev 6
9
Mark the DATE on this file for future reference.
5.2.3
C-arm brake adjustment
Check that the brake has 170 N (38 lbs) of holding force measured at the
cassette holder; if not, open the side covers of the carriage and tighten the
nyloc nuts (3).
If the brake does not release properly, open the nuts (1) and adjust the
space (4) with the screws (2). The space should be about 0.1 to 0.4 mm.
CAUTION!
Make only minor adjustments of the nyloc nuts - measured by the flats of
the nuts.
Date:
Time:
Film type:
Water temp:
Dev temp:
Gross fog:
Speed index:
Contrast
index:
Date:
Time:
Film type:
Water temp:
Dev temp:
Gross fog:
Speed index:
Contrast
index:
Date:
Time:
Film type:
Water temp:
Dev temp:
Gross fog:
Speed index:
Contrast
index:
Table 5.5 Film processor evaluation
Содержание Alpha RT 32551
Страница 1: ...GE Healthcare Alpha RT Installation Instructions 32551 IMG rev 6 0459 ...
Страница 2: ......
Страница 4: ......
Страница 42: ...3 Installation 36 GE Healthcare 32551 IMG rev 6 ...
Страница 51: ...4 Alpha RT softstart procedure 32551 IMG rev 6 GE Healthcare 45 Fig 4 10 Inverter board ...
Страница 52: ...4 Alpha RT softstart procedure 46 GE Healthcare 32551 IMG rev 6 ...
Страница 71: ...5 Setups alignments and adjustments 32551 IMG rev 6 GE Healthcare 65 Fig 5 2 Magnetic brake ...
Страница 94: ...5 Setups alignments and adjustments 88 GE Healthcare 32551 IMG rev 6 ...
Страница 96: ...6 Alpha RT installation step by step 90 GE Healthcare 32551 IMG rev 6 ...
Страница 97: ......
Страница 98: ......