5 Setups, alignments and adjustments
32551-IMG rev 6
GE Healthcare
53
5.1.4
Service menu program #53 PAr
1
Select the program number 53 (PAr) with kV+ button.
2
Press AUTOEXP to enter the PAr program.
#52SUP/i The density display shows i. In this mode the RS 232 link can
be programmed.
0 = no connections
1 = printer or Alpha ID II connected
2 = PC connected (ONLY WITH Alpha ID [Dataview])
3 = Delta 16 mode
4 = Delta 16 / Alpha ID (Dataview), requires a RS-
switcher board
Press
kV+
#52SUP/F The density display shows
F
. This mode determines which
filter type is selected in the power up. Select the mode using
the
density-
and
keys.
0 = selects automatically the rhodium filter
1 = selects automatically the molybdenum filter
2 = selects the previously used filter
3 = selects the auto-filter mode.
Use
mAs-
and
mAs+
keys to determine the
compression thickness (0 - 95 mm in kV display)
where the filter changes.
Press
kV+
Press
Autoexp
to leave the SUP program and to enter the
special program menu.
#53PAr/S The density display shows
S
, which means the
Serial
Number of the Unit
. Check that it equals the serial number
on the label under the image receptor support. If not, correct
using the
density-
and
buttons.
Press
kV+
Содержание Alpha RT 32551
Страница 1: ...GE Healthcare Alpha RT Installation Instructions 32551 IMG rev 6 0459 ...
Страница 2: ......
Страница 4: ......
Страница 42: ...3 Installation 36 GE Healthcare 32551 IMG rev 6 ...
Страница 51: ...4 Alpha RT softstart procedure 32551 IMG rev 6 GE Healthcare 45 Fig 4 10 Inverter board ...
Страница 52: ...4 Alpha RT softstart procedure 46 GE Healthcare 32551 IMG rev 6 ...
Страница 71: ...5 Setups alignments and adjustments 32551 IMG rev 6 GE Healthcare 65 Fig 5 2 Magnetic brake ...
Страница 94: ...5 Setups alignments and adjustments 88 GE Healthcare 32551 IMG rev 6 ...
Страница 96: ...6 Alpha RT installation step by step 90 GE Healthcare 32551 IMG rev 6 ...
Страница 97: ......
Страница 98: ......