9.9.5
Manual instructions - I/O-box
The table below describes the I/O-box related manual instructions available for ÄKTA pure.
Description
Instruction name
Auto zero analog in X sets the value of the analog signal in the
analog port number X to 0 mV.
Auto zero analog in X
Reset auto zero analog in X sets the signal in analog port number
X to its current value, i.e. the actual voltage in the analog port
number X.
Reset auto zero analog in X
Noise reduction analog in X filters the noise in the analog signal
in port number X.
Noise reduction analog in X
Digital out X sets the value of the signal sent out by digital port
number X to either 0 or 1. The default value is 1.
Digital out X
Pulse digital out X generates a pulsed signal in digital port number
X. The signal changes from the initial state (0 or 1) to the opposite
state and returns to the initial state after the defined length of
time.
Pulse digital out X
Configure analog out X enables the user to send one of the pre-
defined signals (UV signal, conductivity, temperature, pH or con-
centration of eluent B) to the analog out port number X, and also
to set the range of that signal.
Configure analog out X
478
ÄKTA pure User Manual 29119969 AB
9 Reference information
9.9 Manual instructions
9.9.5 Manual instructions - I/O-box