9.9.1
Manual instructions - Pumps
The table below describes the pump related manual instructions available for ÄKTA pure.
Description
Instruction name
System flow defines the system flow rate.
System flow
Flow rate can be set either as volumetric or as linear flow. A col-
umn type must be selected before using linear flow.
Sample flow defines the system flow rate. Flow rate can be set
either as volumetric or as linear flow. A column type must be se-
lected before using linear flow.
Sample flow
Gradient sets a gradient (linear or stepwise) using the system
pumps A and B.
Note:
Set gradient length value to 0 to perform a step gradient.
Gradient
Pump A wash is used to change buffers in the inlet tubing, pump
and mixer.
Note:
• Pressing End during Pump wash will terminate both the wash
and the run immediately.
Pressing Continue during Pump wash will terminate the wash
and the run will continue from the point at which the pump
wash instruction was executed.
• An instruction issued when a Pump wash is in progress will not
be executed until the wash is completely finished and all valves
have turned back to the previous positions.
• Pump wash cannot be executed when the system is in state
HOLD.
Pump A wash
ÄKTA pure User Manual 29119969 AB
467
9 Reference information
9.9 Manual instructions
9.9.1 Manual instructions - Pumps