9.2 Module specifications
Introduction
This section specifies the operating data of the components in ÄKTA pure. For general data for the
system see System specifications, on page 411.
System pumps
Data
Parameter
Piston pump, metering type
Pump type
ÄKTA pure 25:
Flow rate range
0.001 to 25 ml/min
(up to 50 ml/min during column packing)
ÄKTA pure 150:
0.01 to 150 ml/min
(up to 300 ml/min during column packing)
ÄKTA pure 25:
Pressure range
0 to 20 MPa (2900 psi)
ÄKTA pure 150:
0 to 5 MPa (725 psi)
ÄKTA pure 25:
Viscosity range
0.35 to 10 cP (5 cP above 12.5 ml/min)
ÄKTA pure 150:
0.35 to 5 cP
ÄKTA pure 25:
Flow rate
specifications
• Accuracy: ± 1.2%
• Precision: RSD < 0.5%
(Conditions: 0.25 to 25 ml/min, < 3 MPa, 0.8 to 2 cP)
ÄKTA pure 150:
• Accuracy: ± 1.5%
• Precision: RSD < 0.5%
(Conditions: 1.0 to 150 ml/min, < 3 MPa, 0.8 to 2 cP)