Inlet valves
The Inlet valves include Inlet valve V9-IA, V9H-IA, V9-IB, V9H-IB, V9-IAB, V9H-IAB, V9-IX and V9H-IX, as
well as Sample inlet valve V9-IS and V9H-IS.
Possible cause and action
Problem
• Hardware error
Restart the instrument with the power switch. If this error is re-
current, generate a System error report and contact Service.
Faulty air sensor in the
valve
pH valve
Possible cause and action
Problem
• The dummy electrode was dry when it was installed in the
valve.
1 Remove the dummy electrode.
2 Wet the dummy electrode properly with distilled water.
3 Insert the dummy electrode into the pH valve.
4 Rotate the dummy electrode before securing it with the nut.
Leaking pH valve
Module Panel
Possible cause and action
Problem
• A Module Panel is missing and the position is left empty
Install the missing Module panel.
• The cable between the Module Panel and the ICU is not con-
nected
Remove the Module Panel and make sure that the cable is con-
nected.
The instrument is unable to
find some of the modules
376
ÄKTA pure User Manual 29119969 AB
8 Troubleshooting
8.4 Troubleshooting: Valves