7.8.4
Replace the O-ring inside the Mixer
Maintenance interval
Replace the O-ring inside the Mixer if it is damaged.
Required material
One of the following O-rings are required:
• O-ring 13.1 x 1.6 mm (for Mixer chambers 0.6, 1.4, and 5 ml)
• O-ring 13.1 x 1.6 mm (highly resistant, for use when the system is exposed to organic solvents or
high concentrations of organic acids such as acetic acid or formic acid for longer periods of time).
• O-ring 22.1 × 1.6 mm (for Mixer chamber 15 ml)
Instruction
Follow the instruction below to replace the O-ring inside the Mixer.
Use a forceps and gloves during the replacement procedure to avoid contaminating the Mixer
components.
Tip:
Action
Step
Loosen the top section of the Mixer.
1
Unscrew the top section of the Mixer and pull apart the Mixer in two halves.
2
Remove the outer locking O-ring from the top section.
3