7.3.3
Clean the Fraction collector F9-C sensors
Maintenance interval
Clean the fraction collector sensors every week, or when required, for example if the fraction collector
fails to read the tray ID or do not collect fractions correctly.
Required material
The following material is required:
• Wash bottle
• Water or 20% ethanol
• Cloth
Instructions
Follow the instructions below to clean the fraction collector diodes. See Section 3.2.2 Fraction collector
F9-C illustrations, on page 93 for the locations of the components of the fraction collector.
Action
Step
In System Control, select Manual:Execute Manual Instructions:Fraction collection:Frac
cleaning position. Click Execute.
1
Result: The Dispenser head moves to cleaning position, and the Instrument display states
System pause.
ÄKTA pure User Manual 29119969 AB
267
7 Maintenance
7.3 Weekly maintenance
7.3.3 Clean the Fraction collector F9-C sensors