Description
Part
Inlet valve
4
System pump B
5
System pump A
6
Pressure monitor
7
Mixer
8
Injection valve
9
Sample loop or Superloop
10
Column valve
11
Column
12
UV monitor
13
Conductivity monitor
14
Flow restrictor
15
Outlet valve
16
Fraction collector
17
Waste
W, W1, W2
Select Mixer chamber
To obtain a homogeneous buffer composition, it is important to use a Mixer chamber suitable for the
flow rate of the method. The tables below show what Mixer chambers to use in ÄKTA pure at different
flow rates.
If the liquids are difficult to mix, use a larger Mixer chamber to achieve optimal mixing. However, note
that a larger Mixer chamber distorts and delays the gradient.
The default mixer size is 1.4 ml for both systems. ÄKTA pure 150 also includes a 5 ml mixer. Other
mixer sizes are available as accessories.
Flow rate [ml/min], Binary gradient
Mixer chamber volume [ml]
ÄKTA pure 150
ÄKTA pure 25
Not recommended
0.1-5
0.6
0.5-15
0.5-15
1.4
2-50
2-25
5
15-150
Not compatible
15
152
ÄKTA pure User Manual 29119969 AB
5 Operation
5.2 Prepare the flow path