4.2.4
Configuration of column valves
System properties
Follow the instruction below to update the system properties.
Action
Step
Open the system properties Edit dialog.
1
Select Valves and pumps from the Component types list.
2
Select components and properties according to the following table:
3
Property
Component selection
Valve
V9-C (5-columns)
Column valve
Column selection valve V9-C
V9H-C (5-columns)
Column valve
Column selection valve V9H-C
V9-Cs (1-column)
Column valve
Column control valve V9-Cs
V9H-Cs (1-column)
Column valve
Column control valve V9H-Cs
System settings
After selecting Column valve V9-Cs, V9H-Cs or no column valve, in UNICORN, the tube length has to
be defined.
The instructions Tubing: Injection valve to column and Tubing: Sample pump to injection valve are
used in calculations of pre-column pressure when Column valve V9-C or V9H-C (5-columns) is NOT
mounted onto the system. In such cases, there is no pre-column pressure sensor present. The estimated
pre-column pressure is calculated using the measured system pressure and the length of the tubing
between the system pressure sensor and the Column.
The instructions are available only when the Column valve V9-C or V9H-C (5-columns) is not selected
in the component list. The instruction Tubing: Sample pump to injection valve is in addition only
available if the sample pump is selected in the list.
Follow the instructions below to set the tube length between the Injection valve and the Column and
between the sample pump and the injection valve.
Action
Step
• In the System Control module, choose System:Connect to Systems or click the Con-
nect to Systems icon.
Result: The Connect to Systems dialog opens.
1
ÄKTA pure User Manual 29119969 AB
135
4 System configuration
4.2 Configure modules
4.2.4 Configuration of column valves