4.2.2
Configuration of Mixer valves
System properties
Follow the instruction below to update the system properties.
Action
Step
Open the system properties Edit dialog.
1
Select Valves and pumps from the Component types list.
2
Select Mixer valve (V9-M) or Mixer valve (V9H-M) in the Component selection list.
3
System settings
The flow rate for Mixer by-pass wash can be set.
Description
Instruction name
Sets the flow rate used during Mixer by-pass wash and defines
wash volumes for mixer by-pass wash options.
Note:
The flow rate should not exceed 10 ml/min if narrow inlet tubing
(i.d. 0.75 mm) is used.
Mixer by-pass wash settings
ÄKTA pure User Manual 29119969 AB
133
4 System configuration
4.2 Configure modules
4.2.2 Configuration of Mixer valves