5.10.1
Set the Cell constant value
Instructions
If the flow cell has been replaced, a new cell constant value has to be set. Follow the in-
structions below to set the Cell constant value.
Action
Step
In the Settings and service screen, tap Next:Next to access the 3
rd
screen.
1
Result: The following screen opens.
Tap Conductivity.
Result: The Conductivity screen opens.
Tap Configuration.
2
Result: The Configuration screen opens.
ÄKTA start Maintenance Manual 29-0603-08 AC
95
5 Service and settings
5.10 Conductivity Monitor
5.10.1 Set the Cell constant value
Содержание akta start
Страница 1: ...ÄKTA start Maintenance Manual Original instructions ...
Страница 2: ...Page intentionally left blank ...
Страница 212: ...Page intentionally left blank ...
Страница 213: ...Page intentionally left blank ...