Troubleshooting Flow chart 2
ERROR 102 UV Monitor Failure
1. Tighten Flow cell and flush it clean with
DM water.
2. Tighten connectors and remove any
strong direct light on the UV Monitor.
3. Tap Continue .
4. Go to Settings and service screen and
perform UV LED calibration.
Switch instrument Off, then On again.
Switch instrument Off, then On again.
1. Tighten and flush the Flow cell.
2. Clean the Flow cell; refer to
ÄKTA start
Operating Instructions.
3. Tap Continue on the error display.
Error message
still displayed?
Switch instrument Off, then On again.
Error message
still displayed?
1. Replace the Flow cell. Follow the Flow cell
replacement procedure.
2. Tap Continue on the error display.
1. Switch instrument Off.
2. Disassemble the UV Monitor.
3. Replace the UV Monitor and Flow cell.
4. Switch instrument On.
Perform UV LED
calibration
Proceed with your work.
Contact GE Service Engineer.
Note:
This Error message is displayed only
during switching on, as part of the power
on self test.
Yes
No
Yes
Yes
Yes
No
No
Yes
Error message
still displayed?
Error message
still displayed?
140
ÄKTA start Maintenance Manual 29-0603-08 AC
6 Troubleshooting
6.8 Troubleshooting flow charts