6.5
Pressure sensor
Error messages
Action
Possible cause
Descrip-
tion
Error code
•
Check tubing and
valves by disconnect-
ing one at a time
starting from the
Fraction collector go-
ing backwards to-
wards the Pump.
When pressure is re-
leased the blockage
has been found, clean
or replace tube to re-
solve the issue.
•
Clean the column/sep-
aration media with
suitable solution (1 M
NaOH) or replace the
column with a new
one.
•
Check the specifica-
tion and pressure of
the columns for cor-
rect flow rate.
•
Blockage in the tubing,
valves or in the col-
umn.
•
Improperly selected
Flow rate.
•
Wrong pressure set-
ting used for chosen
application.
Overpres-
sure
501
•
Perform Zero offset.
•
Contact a GE Service
Engineer.
The sensor is not connect-
ed or not calibrated.
Pressure
sensor fail-
ure
502
ÄKTA start Maintenance Manual 29-0603-08 AC
135
6 Troubleshooting
6.5 Pressure sensor