![GE AKTA ready Скачать руководство пользователя страница 85](http://html.mh-extra.com/html/ge/akta-ready/akta-ready_operating-instructions-manual_128010085.webp)
Open all safety locks on pinch
valves
The safety locks keep the pinch valves locked in an open position, thereby preventing
injuries due to, for instance, a faulty compressed air supply. To open them:
Action
Step
Start the installation by turning
the Installation switch (in the
upper, right corner of the sys-
tem cabinet) to its left position,
FLOW KIT INSTALL. UNICORN
must be in END mode.
1
Open the safety locks on all
pinch valves. The valve safety
locks are opened by turning the
black handles 180 degrees.
There are 19 valves in total.
2
ÄKTA ready Operating Instructions 28960345 AD
85
6 Flow kit installation
6.3 Mounting the flow kit