Computer
Requirements
WARNING
Installing the computer. The computer must be installed and used
according to the instructions provided by the manufacturer of the
computer.
UNICORN is the controlling software for ÄKTA ready. It is recommended that the computer
used is dedicated to UNICORN and related software. Due to the risk for computer viruses,
it is also recommended that the computer is not directly connected to Internet.
See the UNICORN manuals for further details about software requirements and installa-
tion.
Instruction
Two network cables are delivered with the system and should be connected according
to the instruction below:
Action
Step
Connect the encapsulated end of the longer (10 m) straight cable to to the
connector (of type RJ45) marked SYSTEM COMPUTER on the back of the
system.
1
Connect the other end of the long cable to:
2
•
the controlling computer if a direct connection is used, or
•
the shorter (3 m) crossover cable, which, in turn, is connected to a switch
or hub (the shorter cable is only used when connecting to the controlling
computer via a network).
In order not to compromise the protection of the cabinet, use only the enclosed
communication cables.
Note:
ÄKTA ready Operating Instructions 28960345 AD
69
4 Installation
4.6 Connections