Check pressure sensor calibration
In the record table below, enter the True pressure value (reference) and the corresponding
Read value (computer run data).
Remark
Read val-
ue
True val-
ue
Check
Pressure
sensor
Calibration at 0 bar g
PT-111
Calibration at 5 bar g
Calibration at 0 bar g
PT-112
Calibration at 5 bar g
Calibration at 0 bar g
PT-113
Calibration at 5 bar g
Temperature sensor check
Verify that the temperature sensor is working and gives a realistic value. This can be
done by filling a plastic test tube with tempered water and holding it in front of the
temperature sensor. Then check that the temperature (Temp_161) in UNICORN is re-
sponding.
Remark
Sign
Component
TE-161
218
ÄKTA ready Operating Instructions 28960345 AD
14 Reference information
14.6 Checklists