12.4
Storage
If the system is not used for a longer time period (more than a few days):
•
Remove the flow kit (see
Section 11.4 After the run, on page 180
).
•
Clean the cabinet (this chapter).
•
If the system is moved, the system power and air supply should be switched off.
Although the system can be stored at temperatures between -25°C and 60°C,
room temperature is preferable.
Note:
Flow kits should not be stored at temperatures below 5°C.
Do not leave a flow kit mounted on the system for an excessively long time. The tubing
walls may stick together where the tubing is squeezed, possibly causing high pressure
alarms when running the system.
NOTICE
Do not use 0.5 M NaOH (or higher) for more than 2 hours. Flush
afterwards with a suitable rinsing liquid to remove the NaOH.
ÄKTA ready Operating Instructions 28960345 AD