10.2.3
Block 2: Auto-zeroing pressure sensors
Action
Step
Follow the general procedure used
for block 1 to create a new block.
Name the block, for instance,
MyAu-
tozeroPressure
, and set its Base
to Time.
1
Add instructions for auto-zeroing of
the pressure sensors by inserting the
instructions AutoZero_PT111, AutoZe-
ro_PT112, and AutoZero_PT113 (in
group Alarms&Mon), at time 0.10.
2
Note:
This requires that the system is filled with air. If it is filled with liquid, the air
trap valve has to be opened (by pressing the AIR TRAP VENT button with
outlet closed and air trap inline).
The MyAutozeroPressure block now
contains three Autozero instructions,
one for each pressure sensor.