Running the component test
A component test takes approximately 10 minutes. The progress of the test can be fol-
lowed in UNICORN, in charts showing the response curves for different sensors.
During the run, two things will happen that will require user attention:
•
A short while after the test has started, a warning message will appear on the screen
saying that UV stray light reduction is activated. This warning message is information
only and requires no action from the user. The test will proceed normally regardless
of if the user closes the message box (by clicking Acknowledge), or not. The warning
message can be disabled, see
Manual calibration of UV stray light, on page 148
•
A few minutes into the test, a message will appear prompting the user to fill the air
trap. This message requires user action: Press the AIR VENT button to open the air
vent valve (HV-301), allowing for evacuation of air from the air trap.
You must respond to the message regarding evacuation of the air trap. If the
air trap is not properly filled, the test may fail.
Note:
Any time during the test run it is possible to manually add run notes that will be included
in the report. The notes dialog box is opened by clicking Documentation on the View
menu in UNICORN.
112
ÄKTA ready Operating Instructions 28960345 AD
8 Component test
8.1 Run a component test