Action
Step
Select System:Calibrate from the UNICORN System control menu.
1
In the Calibration dialog, choose Conductivity monitor - factory calibration
in the Monitor to calibrate list.
2
Click Restore.
3
Result: The conductivity cell constant is restored to the factory default value.
The conductivity cell constant is printed on the packaging of the Conductiv-
ity monitor.
Conductivity values of 1.00 M
NaCl
The graph below shows the conductivity value for 1.00 M NaCl solution as a function of
temperature.
ÄKTA pilot 600 User Manual 29274559 AA
107
5 Maintenance
5.4 Calibration procedures
5.4.2 Calibrate the conductivity monitor