![GE AKTA pilot 600 Скачать руководство пользователя страница 101](http://html2.mh-extra.com/html/ge/akta-pilot-600/akta-pilot-600_user-manual_4072291101.webp)
5.4.1
Reset the pressure sensors
Introduction
The ÄKTA pilot 600 system is fitted with three pressure sensors, for monitoring system
pressure, pre-column pressure and post-column pressure.
The pressure sensors should be reset at intervals by setting the reading for atmospheric
pressure to zero in UNICORN software. The pressure sensors themselves are not acces-
sible to the user.
This section gives instructions for resetting the pressure sensors.
Maintenance interval
The pressure sensors should be tested every week, or when the the ambient temperature
has changed by more than 5°C. Reset the pressure sensors if the atmospheric pressure
reading is outside the range ± 0.02 MPa (0.2 bar, 2.9 psi).
Tubing connections
Resetting pressure sensors involves disconnecting tubing from the appropriate module
according to the table below.
Take care to avoid spillage when disconnecting tubing.
Note:
Disconnect tubing
Placing
Monitored pres-
sure
Flow restrictor outlet
At the outlet from the flow
restrictor
System pressure
ÄKTA pilot 600 User Manual 29274559 AA
101
5 Maintenance
5.4 Calibration procedures
5.4.1 Reset the pressure sensors
Содержание AKTA pilot 600
Страница 1: ...KTA pilot 600 User Manual...