Possible cause and action
Problem
•
The air intake at the rear or on the right side of the
instrument is covered
Make sure that none of the air intakes on the instru-
ment are covered.
•
Hot surroundings
Decrease the room temperature. Maximum operating
temperature is 35°C.
•
Hardware error
Switch off the instrument and wait until the tempera-
ture has decreased. Restart the instrument. If this error
is recurrent, generate a System error report and con-
tact Service.
The internal instru-
ment temperature is
too high
External air sensors
Possible cause and action
Problem
•
The cable between the external air sensor and the
ICU is not connected
Check the back of the instrument and make sure that
the cable is connected. See illustration below.
•
The external air sensor has not been selected in
UNICORN
Select the external air sensor in UNICORN. Refer to the
installation instructions delivered together with the
external air sensor.
The external air sensor
is not found by the in-
strument
•
One of the connections is not tight enough
Tighten the connectors.
Air is introduced into
the flow path
ÄKTA avant User Manual 29035184 AE
409
8 Troubleshooting
8.7 Troubleshooting: Other components
Содержание AKTA avant