RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUAL
GE MEDICAL SYSTEMS
P9030GX
INSTALLATION AND REPLACEMENT
3–20
REV 21
3–7–5 ECG Leakage Current Test (Continued)
Data Sheet for ECG Leakage Current
Follow the foregoing test procedure. The test passes when all readings measure less than the value shown in the
table below (TABLE 3–8).
TABLE 3–8
MAXIMUM ALLOWANCE LIMIT FOR ECG LEAKAGE CURRENT
Patient Lead to Ground Leakage Current Test
and
Patient Lead to Lead Leakage Current Test
AC POWER
SOURCE
115V
20
µ
A
220/240V
5mA
GROUND
GROUND
OPEN
CLOSED
115V
10
µ
A
10
µ
A
220/240V
500
µ
A
10
µ
A
MAXIMUM
ALLOWANCE LIMIT
AC POWER
SOURCE
MAXIMUM
ALLOWANCE LIMIT
Patient Lead Isolation Current Test
TABLE 3–9
DATA SHEET FOR ECG LEAKAGE CURRENT
ON
NORM
CLOSED
ON
REVERSE
CLOSED
ON
NORM
OPEN
ON
REVERSE
OPEN
OFF
NORM
CLOSED
OFF
REVERSE
CLOSED
OFF
NORM
OPEN
OFF
REVERSE
OPEN
ECG
Tester
Tester
Tester Lead Selector
Power
Polarity
Neutral
Switch
Switch
RL
RA
LA
LL
C
Patient Lead Isolation Current Test
ON
NORM
CLOSED
ON
REVERSE
CLOSED
ON
NORM
OPEN
ON
REVERSE
OPEN
OFF
NORM
CLOSED
OFF
REVERSE
CLOSED
OFF
NORM
OPEN
OFF
REVERSE
OPEN
ECG
Tester
Tester
Tester Lead Selector
Power
Polarity
Neutral
Switch
Switch
RL
RA
LA
LL
C
Patient Lead to Lead Leakage Current Test
ON
NORM
CLOSED
ON
REVERSE
CLOSED
ON
NORM
OPEN
ON
REVERSE
OPEN
OFF
NORM
CLOSED
OFF
REVERSE
CLOSED
OFF
NORM
OPEN
OFF
REVERSE
OPEN
ECG
Tester
Tester
Tester Lead Selector
Power
Polarity
Neutral
Switch
Switch
RL
RA
LA
LL
C
Patient Lead to Ground Leakage Current Test