GE
LOGIQ F S
ERIES
D
IRECTION
5446617-100, R
EVISION
10
B
ASIC
S
ERVICE
M
ANUAL
Chapter 4 - Functional Checks
4 - 33
Section 4-4
Software Configuration Checks
Section 4-5
Peripheral Checks
Check that peripherals work as described below:
Table 4-6
Software Configuration Checks
Step
Task to do
Expected Result(s)
1.
Check Date and Time setting
Date and Time are correct
2.
Check that Location (Hospital Name) is correct
Location Name is correct
3.
Check Language settings
Desired Language is displayed
4.
Check assignment of Printer Keys
The default function for Store and Print Keys are Store
(store image), Print (print). Store and Print Keys can also
be assigned as desired by the customer
5.
Check that all of the customer’s options are set up correct All authorized functions are enabled
Table 4-7
Peripheral Checks
Step
Task to do
Expected Result(s)
1.
Press (
FREEZE
)
Stop image acquisition.
2.
Press (
) on the Control Panel
The image displayed on the screen is printed on B&W printer.
3.
Connect with Foot Switch on USB port and
press once.
To start image acquisition (the same function as (
FREEZE
) key).
Содержание 5478035
Страница 1: ...GE LOGIQ F Series Basic Service Manual Part Number 5446617 100 Revision 10 ...
Страница 2: ......
Страница 3: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL i Important Precautions ...
Страница 4: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL ii ...
Страница 5: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL iii ...
Страница 6: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL iv ...
Страница 7: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL v ...
Страница 8: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL vi ...
Страница 9: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL vii ...
Страница 10: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL viii ...
Страница 11: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL ix ...
Страница 12: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL x ...
Страница 14: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL xii ...
Страница 167: ... 2013 2017 General Electric Company GE GEMedical Systems Ultrasound 9900 Innovation Drive Wawautosa Wisconsin 53226 USA ...