GE
LOGIQ F S
ERIES
D
IRECTION
5446617-100, R
EVISION
10
B
ASIC
S
ERVICE
M
ANUAL
4 - 24
Section 4-3 - General Procedure
4-3-10-3
Worksheets
Measurement/Calculation worksheets are available to display and edit measurements and calculations.
There are generic worksheets as well as Application specific worksheets. The worksheets are selected
from the Measurement Touch Panel.
4-3-10-4
Report Pages
Measurements/Calculations that are included on the worksheet can also be displayed on Report Pages.
Report Pages can be customized to meet the appropriate needs of the user.
4-3-11
Probe/Connectors Usage
4-3-11-1
Connecting a probe
1.) Place the probe's carrying case on a stable surface and open the case.
2.) Carefully remove the probe and unwrap the probe cable.
3.) DO NOT allow the probe head to hang free. Impact to the probe head could result in irreparable
damage.
4.) Align the connector with the probe port and carefully push into place.
5.) Lock the probe.
6.) Carefully position the probe cord so it is free to move and is not resting on the floor.
4-3-11-2
Activating the probe
The probe activates in the currently-selected operating mode. The probe's default settings for the mode
and selected exam are used automatically.
4-3-11-3
Deactivating the probe
1.) Press the
Freeze
key.
2.) Gently wipe the excess gel from the face of the probe. (Refer to the Basic User Manual for
complete probe cleaning instructions.)
3.) Carefully slide the probe around the right side of the keyboard, toward the probe holder. Ensure
that the probe is placed gently in the probe holder.
4-3-11-4
Disconnecting the probe
Probes can be disconnected at any time. However, the probe should not be selected as the active
probe.
1.) Unlock the probe.
2.) Pull the probe and connector straight out of the probe port.
3.) Carefully slide the probe and connector away from the probe port and around the right side of the
keyboard.
4.) Ensure the cable is free.
5.) Be sure that the probe head is clean before placing the probe in its storage box.
WARNING
Take the following precautions with the probe cables: Do not bend, be sure to keep
probe cables free from the wheels.
Содержание 5478035
Страница 1: ...GE LOGIQ F Series Basic Service Manual Part Number 5446617 100 Revision 10 ...
Страница 2: ......
Страница 3: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL i Important Precautions ...
Страница 4: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL ii ...
Страница 5: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL iii ...
Страница 6: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL iv ...
Страница 7: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL v ...
Страница 8: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL vi ...
Страница 9: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL vii ...
Страница 10: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL viii ...
Страница 11: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL ix ...
Страница 12: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL x ...
Страница 14: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL xii ...
Страница 167: ... 2013 2017 General Electric Company GE GEMedical Systems Ultrasound 9900 Innovation Drive Wawautosa Wisconsin 53226 USA ...