8-40
Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
8-3-10-2
Computer Name and Connection Settings
1.) Touch
Utility > Connectivity.
2.) Click on
TCP/IP.
3.) Type
Computer Name.
4.) Verify that
Enable DHCP
has no check mark. If checked, remove the mark.
5.) Check if
Computer name, IP Address, Subnet Mask, and Default Gateway
are proper ones
which you wrote down in section
.
6.) Click on
OK
for confirmation dialog box.
NOTICE
The computer name is unique. Check the scanner serial Number printed on the label located at the rear
lower side of the scanner. When 123456YM1 (for example) is printed,
LS6-123456YM1
must be
entered as a computer name.
NOTICE
If the “Save settings” is NOT performed, you can NOT enter the new patient screen!!
Figure 8-158 TCP/IP
Figure 8-159 Confirmation Dialog Box