Chapter A6: Image Management
Definium AMX 700 X-Ray System
5161515-1EN
Rev. 6 (10 February 2008)
A6-12
Operator Manual
© 2008 General Electric Company. All rights reserved.
Copy Exams to a CD
If your unit has an external CD drive, exams may be copied to a CD for archiving purposes, to send to a
location that is not within the network, or to include with a patient’s medical records.
NOTE:
The CDs used for copying images must be recordable. That is, the CD should be labeled “CD-R”
(recordable).
DO NOT
use re-writable (CD-RW) disks. The system cannot write to this type of disk.
NOTE:
You cannot copy exams to a CD that already has exams or other information saved on it. You will
receive a message, “CD is not blank. Please insert a blank CD to proceed” when you attempt to
copy images to the CD. Always use a new, blank CD.
1. Open the CD tray. (Select Eject from the [CD1] drop-down list.)
NOTE:
Pressing the Eject button on the workstation PC does not open the CD tray. When working on the
Image Management screen, always use the Eject option from the [CD1] drop-down list (Figure A6-
6).
Figure A6-6
CD 1 button drop down list
2. Insert a blank CD into the tray.
3. Press Source [CD1] to close the CD tray.
4. Select the exams to copy.
5. Click the
Destination
[CD1].
The CD Write screen appears.
[DESELECT ALL] unchecks all exams on the list.
[STORE OPTIMALLY] determines which exams to copy to best utilize the space on the CD. It will
automatically uncheck exams that cannot fit on the CD.
[WRITE] begins the copying process.
[CANCEL] closes the screen and returns you to the Image Management screen.