Equipment overview:
Theory of operation
2000966-542D
Dash 3000/4000/5000
2-43
Patient data storage
Static RAM backed up by a lithium battery soldered onto the processor/
power management PCB stores patient data.
Time and date
A patient monitor on the network maintains time and date for at least
five years. Time is accurate to within one second in ten hours. The clock
is synchronized with a time master on the optional CARESCAPE
Network.
Calibration data
Calibration factors for NBP and CO
2
are stored in non-volatile memory
on the DAS board.
Calibration factors for the analog output signals are stored in non-
volatile memory on the processor/power management PCB.
Error log
Input and output errors are retained in static RAM backed up by a
lithium battery soldered onto the processor/power management PCB.
Contents are retained for at least five years, provided that you exercise
caution when you handle the board to prevent inadvertently discharging
the battery (e.g. when you ship the board for problem diagnosis).
Optional thermal printer
The patient monitor uses the same 50-mm thermal printer module that
is used in the PRN 50 stand-alone printer. It prints up to four waveforms
at chart speeds ranging from 0.1 to 50 mm/s. The printer software is
loaded independently from the patient monitor’s software.
In the patient monitor, the printer module limits its current consumption
to stay within its allocated system power budget.
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......