Functional and electrical safety checks:
Functional Checkout procedures
8-22
Dash 3000/4000/5000
2000966-542D
ECG tests
Equipment
Use the following equipment for these tests:
A multiparameter patient simulator.
ECG patient cable
ECG leadwire set
5 leadwire ECG test
Perform this test if the patient monitor acquires 5 leadwire ECG data.
Connections
1. Attach the ECG patient cable and ECG leadwire set to the
ECG/
RESP
connector on the patient monitor and connect the leadwires to
the patient simulator.
Patient Simulator Configuration
2. Set up the patient simulator as follows:
Heart rate – 80 bpm.
Heart rate amplitude – 1.0 mV.
5-leadwire ECG patient cable properly attached.
3. Admit the patient to the monitor.
Normal Sinus Rhythm Procedures
1. Observe the following:
ECG lead II is displayed and is noise-free
Heart rate of 80 ±1 bpm is displayed
With QRS tones enabled, an audible tone sounds with each
Rwave (QRS complex)
2. Verify ECG leads I, II, III, aVL, aVF, and aVR are available to view
and are noise-free.
Pacemaker Detection Procedures
1. Select
DETECT PACE
and set to
PACE 2
.
2. Select a pacemaker pulse on the simulator.
3. Observe the following while you view ECG leads I, II, III, aVL, aVF,
and aVR:
a “P” appears above the PVC count indicating pacemaker pulse
detection is enabled, and
the heart rate still reads 80 ±1 bpm.
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......