Functional and electrical safety checks:
Electrical safety tests
8-12
Dash 3000/4000/5000
2000966-542D
Patient (source) leakage current test
This procedure only applies to Class I (grounded/earthed) equipment,
and measures the leakage current from the ECG/RESP connector or the
SPO2 connector of the device to ground.
NOTE
*The MD represents the network and voltage measuring instrument
and its frequency characteristics per IEC 60601-1.
The patient connector test body shorts all signals in the connector
together. Refer to the instructions contained with the safety analyzer to
perform this test.
1. Connect the ECG/RESP Test Body to the green connector of the
device under test.
2. Configure leakage tester as follows:
Polarity – NORMAL
Neutral – CLOSED
GND (Earth) – CLOSED
3. Power on Device under test.
4. Read and record the current leakage indicated on the tester.
5. Change leakage tester switches to:
Polarity – NORMAL
Neutral – OPEN
GND (Earth) – CLOSED
6. Read and record the current leakage indicated on the tester.
7. Change leakage tester switches to:
Polarity – NORMAL
502B
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......