Troubleshooting:
Review errors
2000966-542D
Dash 3000/4000/5000
6-13
Review errors
This procedure describes how to review the error logs of a patient
monitor. If the CARESCAPE Network software option is enabled, the
error logs may also be transferred over the network to a central station
and copied onto diskette for further review or sent to GE personnel for
review. The transferring procedure is in
WARNING
This procedure is intended for use by service personnel
with advanced troubleshooting skills.
Some of the information recorded in the error logs is useful for field
troubleshooting. The details included here serve as an introduction to the
error logs and provide basic information about what you can learn from
them.
1. Access Review Errors starting from the Main Menu. Select
MORE
MENUS
>
MONITOR SETUP
>
SERVICE MODE
.
2. Enter password using the
Trim Knob
control to select the day and
month from patient monitor screen with leading zeros. (e.g. July 4 =
0407)
3. Select
REVIEW ERRORS
.
The menu provides four error log choices, two for viewing error logs and
two for clearing the error logs.
View output or input errors
1. Select
VIEW OUTPUT ERRORS
or
VIEW INPUT ERRORS
to
view one error in the log of errors.
The error log in a patient monitor holds errors that can be accessed
with the
NEXT
or
PREVIOUS
command. The errors display one
error at a time in the corner of the screen. Watch the error number
category to keep track of which error you are viewing.
The
VIEW OUTPUT ERRORS
provides a list of output software
errors; the
VIEW INPUT ERRORS
provides a list of input software
errors.
2. If you have the writer option you can print an error. Press
or
Graph Go/Stop
.
3. To clear all the errors in the error log, select
CLEAR OUTPUT
ERRORS
or
CLEAR INPUT ERRORS
menu option. Be aware that
once the clear menu option is executed, all selected errors in memory
are erased.
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......