Measurement Formulas
LOGIQ 7/LOGIQ 7 Pro Advanced Reference Manual
2
-
23
Direction 5116269-100 Rev. 2
Formulas–Vascular
Table 2-10: Vascular Calculation Formulas
Calc
Mnemonic
Calc Name
Input
Measurements
Formula
RT ECA
Right External Carotid
Artery Velocity
one Doppler blood flow
peak velocity
RT ECA=v1
[cm/s or m/s]
RT CCA
Right Common Carotid
Artery Velocity
one Doppler blood flow
peak velocity
RT CCA=v1
[cm/s or m/s]
RT BIFURC
Right Carotid Bifurcation
Velocity
one Doppler blood flow
peak velocity
RT BIFURC=v1
[cm/s or m/s]
RT ICA
Right Internal Carotid
Artery Velocity
one Doppler blood flow
peak velocity
RT ICA=v1
[cm/s or m/s]
RT ICA/CCA
Right Internal Carotid
Artery Velocity/Common
Carotid Artery Velocity
Ratio
two Doppler blood flow
peak velocities
RT ICA/CCA=V
ICA
/V
CCA
LT ECA,
LT CCA,
LT BIFURC,
LT ICA,
LT ICA/CCA
Same as above, for Left
Carotid Artery
Same as above
Same as above
A/B Ratio
Velocities Ratio
two Doppler blood flow
peak velocities
A/B=V
1
/V
2
% Stenosis
Stenosis Ratio
two areas (by ellipse,
trace,
circle or distance)
% Stenosis=[1-(A
residual
/
A
lumen
)]x100
S/D Ratio
Systolic Velocity/Diastolic
Velocities Ratio
two Doppler blood flow
peak velocities
S/D=V
systole
/V
diastole
a
PI
Pulsatility Index
two Doppler blood flow
peak velocities and
TAMAX
PI=(V
max
-V
diastole
)/TAMAX
RI
Resistivity Index
two Doppler blood flow
peak velocities
RI=(V
max
-V
diastole
)/ V
max
or
RI=(V
max
-V
min
)/V
max
HR
Heart Rate (beats/minute)
one 2 beat time interval
(measured manually or
automatically)
HR[BPM]=120[sec]/
2 beat time[sec]
Vol. Flow
Volume Flow
dTamean*dDiam*dDiam*0.785
or
dTamax*dDiam*dDiam*0.785
a.V
diastole
= V
min
or V
end-diastole
(depends on preset selection)
Содержание LOGIQ 7
Страница 100: ...Acoustic Output 1 94 LOGIQ 7 LOGIQ 7 Pro Advanced Reference Manual Direction 5116269 100 Rev 2 ...
Страница 150: ...Measurement Formulas 2 50 LOGIQ 7 LOGIQ 7 Pro Advanced Reference Manual Direction 5116269 100 Rev 2 ...
Страница 240: ...OB Tables 3 90 LOGIQ 7 LOGIQ 7 Pro Advanced Reference Manual Direction 5116269 100 Rev 2 ...
Страница 244: ...Index 4 LOGIQ 7 LOGIQ 7 Pro Advanced Reference Manual Direction 5116269 100 Rev 2 ...